Introduction

Look at how the different components are adequately placed inside the MedEasy case and beware of the do's and don'ts.

  1. There is a total of three major components inside the MedEasy case: Raspberry Pi in the blue box.
    • There is a total of three major components inside the MedEasy case:

    • Raspberry Pi in the blue box.

    • Router in the green box.

    • Powerbank in the red box

    • All of them must be carefully secured in the MedEasy case as shown in the image.

  2. The iris scanner can be used with all patients except those suffering from blindness or Amblyopia (Lazy Eye). Patients should remove their spectacles and contact lenses prior to scanning. ALWAYS SCAN THE PATIENT'S RIGHT EYE. Only scan the left eye when the patient's right eye has a disease or there's no right eye.
    • The iris scanner can be used with all patients except those suffering from blindness or Amblyopia (Lazy Eye).

    • Patients should remove their spectacles and contact lenses prior to scanning.

    • ALWAYS SCAN THE PATIENT'S RIGHT EYE. Only scan the left eye when the patient's right eye has a disease or there's no right eye.

    • If you do scan the left eye, you have you have to record this in triage remarks.

    • Always clean the camera (mirror-looking surface) of the iris scanner with an optic cloth both when dirty and when dismantling the equipment prior to leaving the slum.

    • DO NOT clean the iris scanner with any type of cleaning solution!

  3. The OTG micro to micro USB cable is used to connect the iris scanner to the smartphone. You may only use the original OTG cable from IriTech (as shown in the picture, characterised by a red section in one end). For smartphones with ports other than micro USB, please use an adapter in order to connect with the iris scanner.
    • The OTG micro to micro USB cable is used to connect the iris scanner to the smartphone.

    • You may only use the original OTG cable from IriTech (as shown in the picture, characterised by a red section in one end).

    • For smartphones with ports other than micro USB, please use an adapter in order to connect with the iris scanner.

  4. The MedEasy kit and the iris scanner contain sensitive electronics. Proper care must be taken to maintain the maximum functioning capacity of the system.
    • The MedEasy kit and the iris scanner contain sensitive electronics. Proper care must be taken to maintain the maximum functioning capacity of the system.

    • Observe the adequate temperature and humidity range during operation and storage:

    • Temperature: -20°C to +60°C (storage); 0°C to +50°C (operation)

    • Humidity: 10% to 90%

  5. Do not: Allow water to entre the MedEasy kit and the iris scanner.
    • Do not:

    • Allow water to entre the MedEasy kit and the iris scanner.

    • Place the system next to heating equipment.

    • Place magnets near the system.

    • Drop the iris scanner or submit it to sudden impact or mechanical stress.

    • Never disassemble the components from the boards or the iris scanner. Only disassemble when highly necessary or as a last resort.